• 05138768560-09029597412
  • zistkankash@gmail.com, info@toosbioresearch.com

سیستم ثبت سیگنال مغزی

سیستم ثبت سیگنال مغزی
دانلود کاتالوگ
Neural DAQ Specifications

Neural DAQ hardware

Neural DAQ software

Electrode

Camera

Adaptor

Subsystems

40kHz per neural channel and auxiliary non-neural channels at 16-bit resolution

A/D input conversion

25usec

Timestamping

100,500,1000,2000,5000,10000

Gain

2 pole Butterworth

Highpass : 10,50,100,400Hz

Lowpass :  0.5,1,5,10,25kHz

Hardware filter

6 pole Butterworth

Highpass : 1 - 700Hz

Lowpass : 0.8 - 20kHz

Software filter

4 channels

Auxiliary non-neural analog inputs

8 channels

Auxiliary digital input

Simulation and online spike sorting

Up to 500 spikes overlay

Feature space includes PC1-PC2,PC1-PC3,PC2-PC3,Valley-Peak

Offline analysis for up to 500 recent spikes includes clustering (up to 4 clusters) and DCI (Desired Characteristics Identification)

Spike sorting

Simulation and online tracking

Rat tracking

.tdms file

Record up to 10GB per file

File type recorded

Chassis dimensions